- 12 -
Calibration & assembly
III. CALIBRATION
There is no external driver necessary, the EDU09 uses the internal Microsoft Windows HID driver,
these will loaded automatic.
• Set RV1 to the middle position.
• Connect the EDU09 oscilloscope to the USB slot of the PC. The red led should light.
• Run the installed software EDU09.EXE
• Ifitisconnectedforthersttime,calibrationprocedurewillstartautomatically.
• If it does not start automatically: In the Options menu select “Calibrate” and then click OK to
start the calibration. Wait until calibration is successful.
Advanced calibration: Finetuningoftheinputampliercircuit
(requires a 1.5V battery)
This calibration is not really necessary, you must only do this, if you want a higher precision of your
measurements
• In the ‘Options’-menu select “Expert Settings”.
• In the ‘View’-menu select “Waveform Parameters...”.
• In the ‘Waveform Parameters’-window select the check box “DC Mean”.
Bolts
(4x)
Nuts
(4x)
Distance bolts
(4x)
I V. ASSEMBLY
Your scope is now ready for use.
Caps
(4x)
25mm M3 Bolt
15mm
M4 spacer
black cap
M3 nut
Mount the cover.
+
1.5V
+
-
+
-
• Measure the output of the battery with a multi-
meter and remember it.
• Connect the battery to the oscilloscope’s input.
• Set Volts/Div. to “0.5V” and click the “Run”
button.
• Adjust trimmer RV1 until the displayed “DC
Mean” value in the ‘Waveform Parameters’-
window corresponds with the measured value.
• Remove battery.